Vista de pagina 45
Help! There is a comet in my computer! 43
Figure 25. Different versions of user interface for adjusting the camera settings. The main settings
are contrast / brightness, offset / gain, exposure time, binning.
When we are adjusting the camera and we have no image histogram available, we need to
make sure that the image that we see displayed on the screen is a replica of the grabbed
image and not changed by any display settings. In particular, we should check which LUT
is used (the LUT for display of actual grey values usually has the name “grey” or
“greyscale”). Another problem is that not only the camera, but also the display has contrast
and brightness settings. To see the actual grabbed image, the brightness and contrast of the
displayed image should be set to “neutral“(Figure 26). Because the way this is achieved
varies in different software packages, we have to check how to set our own software
correctly.
For accurate recording of comet images, any automatic corrections of the image that a
camera may feature have to be switched off. For example, the setting “gamma”, if
switched on (so that the gamma factor is not 1), distorts camera response away from
linearity.
Once the camera options are adjusted, it is best not to change them throughout the
experiment.